Upon powering up the returned autopulse platform during preventive maintenance, system error, out of service, revert to manual cpr error message was displayed.The root cause was due to the defective processor board.Unable to download archive data.The autopulse platform is a reusable device and was manufactured in october 2011 and is 6 years old.It has exceeded its expected service life of 5 years.As part of routine service during testing, the platform was examined and front enclosure was observed damaged.The front enclosure was replaced.Following the service and repair, the autopulse was tested functionally and passed successfully with no issue or faults observed.
|
The autopulse platform (sn (b)(4)) was returned for preventive maintenance.As part of routine service, the device was tested and during initial power-up, displayed system error, out of service, revert to manual cpr" error message.
|