US SURGICAL PUERTO RICO ENDO GIA ULTRA; STAPLE, IMPLANTABLE Back to Search Results
Model Number EGIAUSHORT
Device Problems Crack (1135); Failure to Fire (2610); Noise, Audible (3273); Firing Problem (4011)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/25/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter: during lobectomy, the surgeon tried to fire the device to pulmonary parenchymal, however at the first squeezing a crack sound was heard and could not fire the device.Replaced by a new handle and a new reload and the firing was completed.The surgeon said pulmonary parenchymal was not thick nor hard.No patient injury.
 
Manufacturer Narrative
Additional information: evaluation summary: post market vigilance (pmv) led an evaluation of one device.Initial visual inspection of the instrument noted no abnormalities.Functionally, the instrument was loaded with an articulating vascular/medium reload.During testing of the instruments, a skip was noted in the units firing stroke after closure of the reload jaws.An access hole was cut into the instrument body for visualization of the firing rack.This examination noted sheared teeth on the firing rack.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of the sheared teeth on the firing rack may occur in any of the following circumstances: 1.Firing over tissue that is beyond the recommended thickness range.2.Firing with an obstacle incorporated in the jaws.In either of these circumstances, it will become increasingly difficult to actuate the firing handle and the instrument return knobs will be difficult to retract.In addition, staples may not form properly, and tissue may not be fully transected.The information booklet which accompanies each product shipment offers the following as a warning and precaution."1.Preoperative radiotherapy may result in changes to tissue.These changes may, for example, cause the tissue thickness to exceed the indicated range thickness for the staple size.Careful consideration should be given to any pre-surgical treatment the patient may have undergone and in corresponding selection of staple size.The root cause of the observed damage was misuse of the product which would have caused or contributed to the reported incident.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO GIA ULTRA
Type of Device
STAPLE, IMPLANTABLE
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer (Section G)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer Contact
lisa hernandez
60 middletown ave
north haven, CT 06473
2034925563
MDR Report Key7682477
MDR Text Key113953364
Report Number2647580-2018-03461
Device Sequence Number1
Product Code GDW
UDI-Device Identifier10884523003536
UDI-Public10884523003536
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K111825
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 10/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEGIAUSHORT
Device Catalogue NumberEGIAUSHORT
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/13/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/27/2018
Initial Date FDA Received07/12/2018
Supplement Dates Manufacturer Received08/01/2018
08/08/2018
Supplement Dates FDA Received08/15/2018
10/03/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-