MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LITTLETON) O-ARM 1000 IMAGING SYSTEM 3RD EDITION; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LITTLETON) O-ARM 1000 IMAGING SYSTEM 3RD EDITION; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE Back to Search Results
Catalog Number BI70000028230
Device Problems Computer Software Problem (1112); No Display/Image (1183)
Patient Problem No Patient Involvement (2645)
Event Date 06/19/2018
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.A medtronic representative went to the site to test the equipment.Testing revealed that the mobile view station (mvs) monitor was displaying a blue screen.Reloaded and upgraded software and the system regained functionality.This software fault is tracked through an internal fault tracking system.Systems checkout was performed and system was functioning as intended.
 
Event Description
Medtronic received information regarding an imaging device being used outside of procedure.It was reported that when the system booted up, the mobile view station (mvs) monitor displayed a blue screen.No patient present.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM 1000 IMAGING SYSTEM 3RD EDITION
Type of Device
IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer Contact
jack edell
navigation customer quality
826 coal creek circle
louisville, CO 80027
MDR Report Key7683506
MDR Text Key113875417
Report Number1723170-2018-03386
Device Sequence Number1
Product Code OXO
Combination Product (y/n)N
Reporter Country CodeCN
PMA/PMN Number
K092564
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 07/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberBI70000028230
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/19/2018
Initial Date FDA Received07/12/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/06/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-