• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC TI DISTAL EXTENSION SIZE 3/220MM RADIUS; PROSTHESIS, RIB REPLACEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC TI DISTAL EXTENSION SIZE 3/220MM RADIUS; PROSTHESIS, RIB REPLACEMENT Back to Search Results
Model Number 04.641.113
Device Problem Break (1069)
Patient Problem No Code Available (3191)
Event Date 06/18/2018
Event Type  Injury  
Manufacturer Narrative
Patient information is unknown.Date of event is unknown.Lot number is unknown.Device malfunctioned intra-operatively and was not implanted / explanted.Complainant part is not expected to be returned for manufacturer review/investigation.Reporter phone number and email address are unknown.Reporter is a synthes employee.Without a lot number the device history records review could not be completed.The investigation could not be completed; no conclusion could be drawn, as no product was received.(b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
(b)(6) reports that during an unknown procedure on an unknown date, the titanium (ti) rod broke when it was bent to be used and thus, was not implanted.A second rod, which was slightly larger, was used in the procedure, and the tail end had to be cut off.Surgical outcome is unknown.There was no surgical delay.There is no reported patient consequence.This report is for a ti distal extension size 3/220mm radius.This is report 1 of 1 for (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Date of event is now known.Additional information provided.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
The event occurred during a lengthening of the vertical expandable prosthetic titanium rib (veptr) rod procedure on (b)(6) 2018.Procedure was completed successfully.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TI DISTAL EXTENSION SIZE 3/220MM RADIUS
Type of Device
PROSTHESIS, RIB REPLACEMENT
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key7689388
MDR Text Key114013981
Report Number2939274-2018-52880
Device Sequence Number1
Product Code MDI
UDI-Device Identifier10705034750484
UDI-Public(01)10705034750484
Combination Product (y/n)N
PMA/PMN Number
K142587
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 06/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number04.641.113
Device Catalogue Number04.641.113
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 06/18/2018
Initial Date FDA Received07/16/2018
Supplement Dates Manufacturer Received08/08/2018
09/11/2018
Supplement Dates FDA Received08/15/2018
09/11/2018
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-