Model Number C12059
Device Problem Material Fragmentation (1261)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/28/2018
Event Type  Injury  
Manufacturer Narrative
The reported event of "the tip of the pressurewire x, wireless separated" could not be confirmed.The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported incident could not be conclusively determined.The pressurewire aeris instructions for use (ifu) cautions that torqueing the pressurewire against resistance or repeated attempts to cross a total vessel occlusion may cause damage and/or fracture, which may lead to a portion of pressurewire separating from the tip.The pressurewire x instructions for use (ifu) states that excessive manipulation of the pressurewire when the sensor element or pressurewire tip is located in sharp bend may cause damage or tip fracture.The pressurewire x instructions for use (ifu) states that torqueing the pressurewire against resistance or repeated attempts to cross a total vessel occlusion may cause damage and/or fracture, which may lead to a portion of pressurewire separating from the tip.
 
Event Description
During a procedure in the left anterior descending artery, the tip of the pressurewire x, wireless separated and remained in the vessel.A snare device was used to successfully retrieve the separated tip from the patient's anatomy.The procedure was successfully completed without ffr.There was no adverse patient effect.There was a delay in the procedure due to snaring the separated tip; however, no significant impact to the patient.The patient age, date of birth, gender, weight and other relevant medical history were not provided due to patient privacy policy.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRESSUREWIRE¿ X, WIRELESS, RX 175CM
Type of Device
TRANSDUCER, PRESSURE, CATHETER TIP
Manufacturer (Section D)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
elizabeth boltz
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key7691405
MDR Text Key114091753
Report Number3008452825-2018-00226
Device Sequence Number1
Product Code DXO
UDI-Device Identifier05415067025715
UDI-Public05415067025715
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161171
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial
Report Date 07/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2019
Device Model NumberC12059
Device Catalogue NumberC12059
Device Lot Number6240084
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/02/2018
Initial Date FDA Received07/16/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/08/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-