• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEXCOM, INC. DEXCOM G6 CONTINUOUS GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEXCOM, INC. DEXCOM G6 CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 9500-45
Device Problem Retraction Problem (1536)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/17/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
Dexcom was made aware on (b)(4) 2018, that on (b)(6) 2018, a needle retraction issue occurred.The sensor was inserted into the abdomen on (b)(6) 2018.No product or data was provided for evaluation.Confirmation of the problem and probable cause could not be determined.No additional event or patient information is available.
 
Manufacturer Narrative
Com-(b)(4).
 
Event Description
A sensor applicator was returned for evaluation.Based on visual inspection, the applicator was fully deployed.The reported event of a needle retraction issue could not be confirmed.A probable cause could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G6 CONTINUOUS GLUCOSE MONITORING SYSTEM
Type of Device
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence dr.
san diego CA 92121
Manufacturer Contact
kipp durbin
6340 sequence dr.
san diego, CA 92121
8582000200
MDR Report Key7695025
MDR Text Key114438803
Report Number3004753838-2018-078365
Device Sequence Number1
Product Code MDS
UDI-Device Identifier00386270000255
UDI-Public00386270000255
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P170031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date12/20/2018
Device Model Number9500-45
Device Catalogue NumberSTS-OR-001
Device Lot Number5238402
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/27/2018
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/17/2018
Initial Date FDA Received07/17/2018
Supplement Dates Manufacturer Received10/13/2018
Supplement Dates FDA Received10/28/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/24/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age57 YR
-
-