Correction: due to an it issue beginning on (b)(6) 2018, previously filed emdrs did not contain required fields.This supplemental emdr is filed to provide the following omitted fields: b.2.Event attributed to: other.D.8.Device single use?: no.H.3.Device returned to manufacture: no.H.6.Investigation: investigation summary: the investigation determined the barcode number (and therefore the barcode itself) for the accu-sed® normal control (level 1) of lot numbers #1823500, #1823500-1, #1823500-2 and #1823500-3 was incorrect for the monitor and bd sedi instruments.This could result in problems to enter the data for this level for these instrument types that are equipped with a qc program.Bd had initiated further investigation relating to this issue through a capa and as a result, corrective actions were implemented.A revised ifu (rev.C) containing the correct barcodes was communicated to all bd customers on (b)(6) 2018.Investigation conclusion: the customer's indicated failure mode for the barcode error was observed by bd.Further investigation activities were conducted through a capa and the most likely root cause was identified.As a result, corrective actions and procedures were implemented to mitigate further occurrences.Root cause description: the root cause was attributed to a manufacturing incident.Bd had initiated further investigation through a capa.The capa had identified the most likely root causes and corrective actions were implemented.Rationale: based on an assessment of severity and frequency, it was determined that a capa was required in order to determine the root cause associated with this issue.The investigation identified root cause(s) for this issue and corrective actions were implemented.
|