• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: US SURGICAL PUERTO RICO ENDO GIA ULTRA; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

US SURGICAL PUERTO RICO ENDO GIA ULTRA; STAPLE, IMPLANTABLE Back to Search Results
Model Number EGIAUSHORT
Device Problems Failure to Advance (2524); Failure to Fire (2610)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/02/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, at the third firing of lobectomy, the surgeon pushed the green button and tried to fire the device to the lung blood vessel, however, the knife did not advance.The procedure was completed with another device.No patient injury.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Additional info: if information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Additional info: evaluation summary: post market vigilance (pmv) led an evaluation of one device.Visual inspection of the returned product noted that the firing knobs were retracted and the articulation lever was in neutral position.The instrument was loaded with a pmv representative reload and applied to test media.All staples were placed and the test media was cleanly transected.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all medtronic quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO GIA ULTRA
Type of Device
STAPLE, IMPLANTABLE
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer (Section G)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer Contact
lisa hernandez
60 middletown ave.
north haven, CT 06473
2034925563
MDR Report Key7699362
MDR Text Key114678861
Report Number2647580-2018-03573
Device Sequence Number1
Product Code GDW
UDI-Device Identifier10884523003536
UDI-Public10884523003536
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K111825
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date 12/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2022
Device Model NumberEGIAUSHORT
Device Catalogue NumberEGIAUSHORT
Device Lot NumberP7H1071X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/30/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/03/2018
Initial Date FDA Received07/18/2018
Supplement Dates Manufacturer Received08/08/2018
11/24/2018
11/24/2018
Supplement Dates FDA Received10/03/2018
12/19/2018
12/19/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/04/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-