Model Number EGIAUSHORT |
Device Problems
Failure to Advance (2524); Failure to Fire (2610)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 07/02/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, at the third firing of lobectomy, the surgeon pushed the green button and tried to fire the device to the lung blood vessel, however, the knife did not advance.The procedure was completed with another device.No patient injury.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Additional info: if information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Additional info: evaluation summary: post market vigilance (pmv) led an evaluation of one device.Visual inspection of the returned product noted that the firing knobs were retracted and the articulation lever was in neutral position.The instrument was loaded with a pmv representative reload and applied to test media.All staples were placed and the test media was cleanly transected.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all medtronic quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.If information is provided in the future, a supplemental report will be issued.
|
|