The patient was undergoing a coil embolization procedure in the hypogastric artery using pod10s.During the procedure, while attempting to advance a pod10 through a lantern delivery microcatheter (lantern), the physician experienced resistance and the pod10 pusher assembly became kinked.Subsequently, the pod10 would not advance smoothly and therefore, was removed.The procedure was completed using another pod10 and the same lantern.There was no report of adverse effect to the patient.
|
Results: the pod10 had bends at approximately 32.0 cm and 44.0 cm from the proximal end of the pusher.Kinks were found approximately 78.5 cm and 119.5 cm from the proximal end of the pusher.A fracture was found approximately 118.0 cm from the proximal end of the pusher.The embolization coil was detached from its pusher assembly inside the introducer sheath.The pod10 was fractured; therefore, functional analysis could not be performed.Conclusions: evaluation of the returned pod10 revealed kink, bend, and fracture damage along the length of the pusher assembly.These damages may have been a result of forcefully advancing the device against resistance.As reported in the complaint, some of the damage occurred outside the patient body post-procedure.The microcatheter identified in the complaint was not returned for evaluation; therefore, the root cause of the resistance experienced could not be determined.Further evaluation revealed the embolization coil was detached.If the pusher assembly becomes fractured it will likely allow the pull wire to retract from the distal detachment tip (ddt), detaching the embolization coil.Penumbra coils are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|