The hill-rom technician found the brake casters needed to be replaced.Per the hill-rom user manual, warning: patients may use the bed for support while entering or exiting; if the unit moves unexpectedly, patient injury could occur.When the unit is unattended, ensure that both brakes are locked.The brakes for the clinitron bed are located at the right, head end and the left, foot end of the unit.To apply the brakes, step on the lower end of the brake lever to lock the wheels.To release the brakes, apply inward pressure to the upper end of the brake lever.This is a rental product hill-rom maintains and performs the preventative maintenance.The technician removed this bed from the account and provided the account with another clinitron bed.The technician replaced this bed's brake casters to resolve the issue.Based on this information, no further action is required.
|