• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. HARMONIC ACE 36CM W ERG HANDLE; INSTRUMENT, ULTRASONIC SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. HARMONIC ACE 36CM W ERG HANDLE; INSTRUMENT, ULTRASONIC SURGICAL Back to Search Results
Catalog Number ACE36E
Device Problems Break (1069); Loose or Intermittent Connection (1371); Noise, Audible (3273)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/03/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # p9480t.The lot history records were reviewed and the manufacturing criteria were met prior to the release of this lot.
 
Event Description
It was reported that during an unknown procedure, solid tone with error code 'tighten assembly' and then the titanium tip broke during the procedure.The broken piece was retrieved by forceps and was discarded.Changed to another device to complete the procedure.There was no report on patient injury.
 
Manufacturer Narrative
(b)(4) investigation summary: the device was returned with the distal tip of the blade broken off and not returned.The remaining blade portion was scratched and had evidence of contact with metal in or out of the operative field.During functional testing on a gen11, an alert screen was displayed.A probable cause for the device to stop activating and the gen11 to display an alert screen is blade damage probable causes of blade damage, including breakage, are external contact during pre-op or general use, blade contact with other devices, contact with staples or clips during the procedure or using any means other than the blade wrench to attach or detach the blade.Once minor blade damage has occurred, subsequent activation may increase the severity of the blade damage.This, in turn, can result in the device failing the pre-run test with the generator and displaying an alert screen.The alert screens that can result may include ¿tighten assembly,¿ ¿blade error detected,¿ or "relax pressure on blade," followed by a ¿replace instrument¿ screen later in the procedure.Continued usage of the damaged blade can result in a broken blade.The batch history record was reviewed and no defects, nc¿s or protocols related to the complaint, were found during the manufacturing process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HARMONIC ACE 36CM W ERG HANDLE
Type of Device
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
MDR Report Key7700891
MDR Text Key114688220
Report Number3005075853-2018-11371
Device Sequence Number1
Product Code LFL
UDI-Device Identifier20705036000935
UDI-Public20705036000935
Combination Product (y/n)N
PMA/PMN Number
K042777
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 07/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Expiration Date10/31/2022
Device Catalogue NumberACE36E
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/19/2018
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/04/2018
Initial Date FDA Received07/18/2018
Supplement Dates Manufacturer Received07/04/2018
Supplement Dates FDA Received07/27/2018
Patient Sequence Number1
Treatment
GENERATOR AND HANDPIECE
-
-