• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HILL-ROM BATESVILLE VERSACARE FRAME; A/C POWERED ADJUSTABLE HOSPITAL BED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HILL-ROM BATESVILLE VERSACARE FRAME; A/C POWERED ADJUSTABLE HOSPITAL BED Back to Search Results
Model Number P3200K000500
Device Problem Positioning Problem (3009)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/05/2018
Event Type  malfunction  
Manufacturer Narrative
The hill-rom technician found the external alarm needed to be replaced.Per the hill-rom service manual, perform annual preventive maintenance procedures to make sure all versacare® bed components are functioning as originally designed.Make sure the brake is not set, and then plug the bed into an applicable power source.A steady alarm comes on.Set the brake.The alarm stops.Repair or replace as necessary.A search of the hill-rom maintenance records did not show hill-rom performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The technician replaced the external alarm to resolve the issue.Based on this information, no further action is required.
 
Event Description
Hill-rom received a report from a hill-rom technician stating the brake not set alarm would not sound.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint #: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSACARE FRAME
Type of Device
A/C POWERED ADJUSTABLE HOSPITAL BED
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
tony werner
1069 state route 46 east
batesville, IN 47006
8129312359
MDR Report Key7702177
MDR Text Key114722880
Report Number1824206-2018-00277
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 07/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberP3200K000500
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/05/2018
Initial Date FDA Received07/19/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/14/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-