Catalog Number PVE35A |
Device Problems
Defective Component (2292); Noise, Audible (3273)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 06/22/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).We did not receive a batch or lot number for the product involved in this complaint.Therefore, we were unable to check manufacturing records for any related non-conformance.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date, a supplemental medwatch will be sent.When the device failed to staple, did it cut the tissue?.
|
|
Event Description
|
It was reported that during a thoracotomy procedure the device made a funny noise on the first firing attempt across the pulmonary vessel.The tried to use the reverse button which did not work.They manually opened the device and it was noticed that although no staples deployed all of the drivers were up.The procedure was completed using a like device.There were no patient consequences reported.
|
|
Manufacturer Narrative
|
(b)(4).Date sent: 8/21/2018.Batch # r5772d.The analysis found that one pve35a device was returned with no apparent damage and with one cartridge reload loaded on the device.The reload was received fully fired.The device was tested for functionality in the straight position with a test cartridge reload and achieved its complete firing sequence without any difficulties.The staple line and cut line were complete and the staples were noted to have the proper b-form shape.The event described could not be confirmed as the device performed without any difficulties noted.The knife reverse button worked properly during testing.The device opened and closed without any difficulties noted.The batch history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the manufacturing process.
|
|
Search Alerts/Recalls
|