• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. ECHELON FLEX POWERED VASCULAR STAPLER WITH ADVANCED PLACEMENT TIP; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. ECHELON FLEX POWERED VASCULAR STAPLER WITH ADVANCED PLACEMENT TIP; STAPLE, IMPLANTABLE Back to Search Results
Catalog Number PVE35A
Device Problems Defective Component (2292); Noise, Audible (3273)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/22/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).We did not receive a batch or lot number for the product involved in this complaint.Therefore, we were unable to check manufacturing records for any related non-conformance.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date, a supplemental medwatch will be sent.When the device failed to staple, did it cut the tissue?.
 
Event Description
It was reported that during a thoracotomy procedure the device made a funny noise on the first firing attempt across the pulmonary vessel.The tried to use the reverse button which did not work.They manually opened the device and it was noticed that although no staples deployed all of the drivers were up.The procedure was completed using a like device.There were no patient consequences reported.
 
Manufacturer Narrative
(b)(4).Date sent: 8/21/2018.Batch # r5772d.The analysis found that one pve35a device was returned with no apparent damage and with one cartridge reload loaded on the device.The reload was received fully fired.The device was tested for functionality in the straight position with a test cartridge reload and achieved its complete firing sequence without any difficulties.The staple line and cut line were complete and the staples were noted to have the proper b-form shape.The event described could not be confirmed as the device performed without any difficulties noted.The knife reverse button worked properly during testing.The device opened and closed without any difficulties noted.The batch history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the manufacturing process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ECHELON FLEX POWERED VASCULAR STAPLER WITH ADVANCED PLACEMENT TIP
Type of Device
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
milton garrett
475 calle c
guaynabo 00969
5133378865
MDR Report Key7704599
MDR Text Key114707257
Report Number3005075853-2018-11425
Device Sequence Number1
Product Code GDW
UDI-Device Identifier20705036014581
UDI-Public20705036014581
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K141952
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Expiration Date12/24/2020
Device Catalogue NumberPVE35A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/26/2018
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/25/2018
Initial Date FDA Received07/19/2018
Supplement Dates Manufacturer Received07/26/2018
Supplement Dates FDA Received08/21/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/24/2018
Is the Device Single Use? Yes
Type of Device Usage N
Patient Sequence Number1
-
-