ETHICON ENDO-SURGERY, LLC. HARMONIC SCALPEL HAND PIECE; INSTRUMENT, ULTRASONIC SURGICAL
|
Back to Search Results |
|
Catalog Number HP054 |
Device Problem
Loss of Power (1475)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 05/14/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Batch # p93a1x.Investigation summary: the device was received with no apparent damage.It was connected to a generator, evaluated with a test instrument and was found to be non-functional.The instrument was disassembled to inspect the internal components.The moisture indicator was positive.The handpiece has a number of seals to prevent fluids from entering the housing."positive moisture indicator¿ describes a condition where water enters the handpiece cavity during the steam sterilization process.The primary path of ingress is the distal seal, this may be caused by a reduction of the compressive force on the distal seal.However, no definitive root cause could be drawn.It is probable that the ingress of moisture affected handpiece functionality.The manufacturing records were reviewed and the manufacturing/packaging criteria were met prior to the release of this batch.
|
|
Event Description
|
It was reported that during a laparoscopic appendectomy procedure, it was deactivating.Generator must be restarted for use, however, it did not work.Another device had to be used to proceed with the surgery.There was no patient consequence.
|
|
Search Alerts/Recalls
|
|
|