• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC XOMED INC. XPS® BLADE; BUR, EAR, NOSE AND THROAT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC XOMED INC. XPS® BLADE; BUR, EAR, NOSE AND THROAT Back to Search Results
Model Number 1884012HR
Device Problems Material Separation (1562); Detachment of Device or Device Component (2907)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 07/02/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
A healthcare provider (hcp) reported via manufacturer representative that during a dacryocystorhinostomy, the inner cylinder of the blade came off.There was no intervention performed or planned.There was no delay in the procedure and there was no patient impact or injury.
 
Manufacturer Narrative
Analysis of the blade found the inner assembly would spin and the middle would index freely by hand.When viewed under magnification, there was damage to the hubs that is consistent with improper loading: dimples on the front hub prior to the locking area caused by the handpiece locking mechanism; locking area deformation caused by the back side of the front collet of the handpiece; and minor deformation of the inner hub chevrons caused by the handpiece drive mechanism.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XPS® BLADE
Type of Device
BUR, EAR, NOSE AND THROAT
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer (Section G)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer Contact
uriza shums
6743 southpoint drive north
jacksonville, FL 32216
9043328405
MDR Report Key7709631
MDR Text Key114940009
Report Number1045254-2018-00307
Device Sequence Number1
Product Code EQJ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 10/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1884012HR
Device Catalogue Number1884012HR
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/23/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/18/2018
Initial Date FDA Received07/23/2018
Supplement Dates Manufacturer Received08/01/2018
08/08/2018
Supplement Dates FDA Received08/27/2018
10/02/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-