Model Number 1884012HR |
Device Problems
Material Separation (1562); Detachment of Device or Device Component (2907)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
Event Date 07/02/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
A healthcare provider (hcp) reported via manufacturer representative that during a dacryocystorhinostomy, the inner cylinder of the blade came off.There was no intervention performed or planned.There was no delay in the procedure and there was no patient impact or injury.
|
|
Manufacturer Narrative
|
Analysis of the blade found the inner assembly would spin and the middle would index freely by hand.When viewed under magnification, there was damage to the hubs that is consistent with improper loading: dimples on the front hub prior to the locking area caused by the handpiece locking mechanism; locking area deformation caused by the back side of the front collet of the handpiece; and minor deformation of the inner hub chevrons caused by the handpiece drive mechanism.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|