U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CORDIS CASHEL 466P306X; CATHETER, INTRAVASCULAR, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CORDIS CASHEL 466P306X; CATHETER, INTRAVASCULAR, DIAGNOSTIC Back to Search Results
Catalog Number 466P306X
Device Problems Fracture (1260); Unintended Movement (3026)
Patient Problem Not Applicable (3189)
Event Date 07/05/2018
Event Type  Injury  
Manufacturer Narrative
Event date is unknown.Implant date is unknown.As reported, the patient underwent placement of a trapease inferior vena cava (ivc) filter.The filter subsequently malfunctioned and caused injury, damage to patient, including, but not limited to: pain and suffering due to fractured filter with strut embedded near heart.The product was not returned for analysis and the sterile lot number has not been provided; therefore, no device analysis nor device history record review could be performed.The inferior vena cava filter is indicated for use in the prevention of recurrent pulmonary embolism (pe) via percutaneous placement in the vena cava for patients in which anticoagulants are contraindicated, anticoagulant therapy for thromboembolic disease has failed, emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced or for chronic, recurrent pulmonary embolism where anticoagulant therapy has failed, or is contraindicated.The purpose of a vena cava filter is to catch thrombus from the lower extremities as it travels along normal blood flow patterns up towards the heart.Due to the nature of the complaint, the reported pain and suffering experienced by the patient could not be confirmed and the exact cause could not be determined.Without procedural films available for review, the reported ¿filter- fractured-separated - in patient¿ and ¿filter- migration¿ could not be confirmed and the exact cause could not be determined.The instructions for use (ifu) states filter fracture is a potential complication of vena cava filters.Anatomic locations that create concentrated stress points from filter deformation (for example, deployment at apex of scoliosis, overlapping of either of the renal ostia, or placement adjacent to a vertebral osteophyte) may contribute to fracture of a particular filter strut.Inferior vena cava (ivc) filter migration is a known potential adverse event associated with all ivc filter implants and is listed in the instruction for use (ifu) as such.Possible causes for filter migration includes mega cava, wire entrapment during central venous catheter placement, ¿sail¿ effect (cranial migration) of large clot burden within the filter, mechanical device failure, and operator error.Physiologic causes of migration may result from temporary dysmorphism of the inferior vena cava including bending, coughing or valsalva maneuvers resulting in dislodgment of the filter.Some studies suggest that strenuous physical activity and increased intra-abdominal pressure can lead to migration of ivc filters.Giving the limited information available at this time, clinical factor contributing to the migration could not be determined.Also, with the limited information available for review at this time, there is nothing to suggest that the reported event is related to the design and/or manufacturing process of the device; therefore no corrective action will be taken.Should additional information become available, the file will be updated accordingly.
 
Event Description
As reported by the legal brief, the patient underwent placement of a trapease inferior vena cava (ivc) filter.The filter subsequently malfunctioned and caused injury, damage to patient, including, but not limited to: pain and suffering due to fractured filter with strut embedded near heart.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
466P306X
Type of Device
CATHETER, INTRAVASCULAR, DIAGNOSTIC
Manufacturer (Section D)
CORDIS CASHEL
cahir road
cashel, co. tipperary
EI 
Manufacturer (Section G)
CORDIS CASHEL
cahir road
cashel, co. tipperary
EI  
Manufacturer Contact
karla castro
14201 nw 60th ave
miami lakes, FL 33014
7863138372
MDR Report Key7714476
MDR Text Key114853462
Report Number1016427-2018-01673
Device Sequence Number1
Product Code DQO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K020316
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number466P306X
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/05/2018
Initial Date FDA Received07/24/2018
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Life Threatening;
-
-