• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - TUNISIA BLOOD RECIPIENT SET; SET, BLOOD TRANSFUSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - TUNISIA BLOOD RECIPIENT SET; SET, BLOOD TRANSFUSION Back to Search Results
Catalog Number FMC9673
Device Problems Material Deformation (2976); No Flow (2991)
Patient Problem No Patient Involvement (2645)
Event Date 06/25/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(6).The device was received for evaluation.Visual inspection revealed that the tubing was kinked just under the chamber.Gravity testing was also performed, and solution was found to not flow through the tube under the chamber due to excess solvent causing blockage at the connection of the chamber and the tubing.The reported no flow condition was verified.The cause of the excess solvent was determined to be an operator error during manufacturing.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a blood administration pump set was unable to flow due to a kink in the tubing.This event occurred during priming.There was no patient involvement.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BLOOD RECIPIENT SET
Type of Device
SET, BLOOD TRANSFUSION
Manufacturer (Section D)
BAXTER HEALTHCARE - TUNISIA
tunis
Manufacturer (Section G)
BAXTER HEALTHCARE - TUNISIA
route de chebbaou
2021oued e
tunis
TS  
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key7714524
MDR Text Key114882693
Report Number1416980-2018-04570
Device Sequence Number1
Product Code BRZ
UDI-Device Identifier05413760170107
UDI-Public(01)05413760170107
Combination Product (y/n)N
Reporter Country CodeSN
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 07/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2022
Device Catalogue NumberFMC9673
Device Lot Number17L16T835
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/17/2018
Initial Date Manufacturer Received 06/25/2018
Initial Date FDA Received07/24/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/28/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-