MAUDE Adverse Event Report: US SURGICAL PUERTO RICO GIA; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

US SURGICAL PUERTO RICO GIA; STAPLE, IMPLANTABLE Back to Search Results
Model Number GIA8038L
Device Problems Failure to Advance (2524); Misfire (2532)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/09/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during a cystectomy/ileal conduit, the surgeon started firing the device; however preset cartridge did not advance on the halfway.Replaced by new cartridge, same event occurred.They replaced the handle and firing was successful.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.Visual inspection of the instrument noted that the center bar had retracted in the retainer.The loading unit displayed a full complement of staples and the interlock was set.No damage was noted to the knife blade.The single used loading unit(sulu) was loaded into a test unit.The test instrument and sulu were applied to the appropriate test media with acceptable results.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of the center bar retraction condition may occur if the loading units are unloaded improperly.After firing, if the firing knob is not fully retracted, difficulty in removal of the loading unit may be experienced and the center bar may fall back into the firing knob retainer.The root cause of the observed damage was misuse of the product which caused or contributed to the reported condition.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GIA
Type of Device
STAPLE, IMPLANTABLE
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer (Section G)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer Contact
lisa hernandez
60 middletown ave
north haven, CT 06473
2034925563
MDR Report Key7718569
MDR Text Key115195261
Report Number2647580-2018-03719
Device Sequence Number1
Product Code GDW
UDI-Device Identifier10884523007251
UDI-Public10884523007251
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K111825
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,user faci
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 10/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2023
Device Model NumberGIA8038L
Device Catalogue NumberGIA8038L
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/10/2018
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 07/10/2018
Initial Date FDA Received07/25/2018
Supplement Dates Manufacturer Received08/08/2018
09/24/2018
Supplement Dates FDA Received10/02/2018
10/17/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/14/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-