EDWARDS LIFESCIENCES DR PRESSURE MONITORING KIT; TRANSDUCER, PRESSURE, CATHETER TIP Back to Search Results
Model Number T100214A
Device Problems Detachment of Device or Device Component (2907); Insufficient Information (3190)
Patient Problem Air Embolism (1697)
Event Date 12/31/2017
Event Type  Injury  
Manufacturer Narrative
The device was not returned for evaluation as it was discarded at the hospital.Without return of the unit, it is not possible to determine if some damage or defect existed on the unit that could have contributed to the event.No corrective actions will be taken at this time.Lot number was not provided; therefore, review of the manufacturing records could not be completed.It is common clinical practice to inspect all products before use.These products are used by highly trained clinicians, experienced in identifying and mitigating any hazards that arise during use.If the pressure tubing becomes detached during use, it will affect the pressure waveform, which will immediately alert the clinician to begin the troubleshooting process.The ifu states: ¿do not allow air bubbles to enter the setup, especially when monitoring arterial pressures.¿ ¿if all air is not removed from the bag, air may be forced into the patient¿s vascular system when the solution is exhausted.¿ ¿significant distortion of the pressure waveform or air emboli can result from air bubbles in the setup.¿ ¿periodically check fluid path for air bubbles.Ensure that connecting lines and stopcocks remain tightly fitted.¿ air emboli is a stated complication in the ifu.Invasive procedures involve some patient risks.Although serious complications are relatively uncommon, the physician is advised to consider the potential benefits in relation to the possible complications.It is unknown whether user or procedural factors played a role in this event.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
 
Event Description
It was reported that during use, the tubing of the disposable pressure transducer (dpt) disconnected at the zero stopcock.The dpt was connected to a cvp line.It was indicated that the patient suffered an air embolism while sitting in a chair.Follow up with hospital indicated that after treatment the patient¿s condition was favorable and no other complications were reported.The device was not available for evaluation since it was discarded.
 
Manufacturer Narrative
Corrected data: f10, h6.Reference capa-20-00141.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRESSURE MONITORING KIT
Type of Device
TRANSDUCER, PRESSURE, CATHETER TIP
Manufacturer (Section D)
EDWARDS LIFESCIENCES DR
parque industrial de itabo
carr. sanchez km 18.5
haina, san cristobal
DR 
MDR Report Key7721043
MDR Text Key115060768
Report Number2015691-2018-03017
Device Sequence Number1
Product Code DXO
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 07/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberT100214A
Device Catalogue NumberT100214A
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 07/03/2018
Initial Date FDA Received07/25/2018
Supplement Dates Manufacturer Received07/23/2020
Supplement Dates FDA Received11/06/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-