Same case as: 2134265-2018-06508, 2134265-2018-06509.It was reported that patient experienced a pericardial effusion.During an atrial fibrillation ablation procedure performed on (b)(6) 2018 in which an intellanav mifi xp ablation catheter, an intellamap orion catheter and a dynamic xt catheter were used, the patient experienced a pericardial effusion.The effusion occurred after a series of 9 rf ablations (following 15 minutes of mapping with the orion catheter) in the anterior wall, close to the left atrial appendage ostium.The company representative present at the procedure did not observe any strange impedance values or value variations and the intellanav mifi xp catheter did not go out of geometry.For the first 8 rf lesions, the generator parameters were: 35w, 30 degrees c, approximately 90 ohms.The 9th lesion generator parameter was 30w, 30 degrees c, approximately 87 ohms.The average impedance drop for the 9 rf ablations was 15 to 25 ohms (largest = -26.4 ohms and the impedance drop for the 9th rf ablation was -20.0 ohms).The starting impedance value was around 105 ohms for every rf application with 115 ohms being the good tissue contact impedance reference.The effusion was noticed immediately by the physician because the patient¿s blood pressure had decreased rapidly.The physician suspected that the last rf ablation was responsible for the effusion.One liter of blood was drained from the patient¿s pericardium and he was transferred to the surgery lab.The patient stabilized after 30 minutes and remained stable after 24 hours.
|