Model Number M00564520
Device Problems Difficult to Remove (1528); Positioning Problem (3009); Activation Failure (3270)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/04/2018
Event Type  Injury  
Manufacturer Narrative
(b)(4).According to the complainant, the suspect device has been disposed and is not available for return.If any further relevant information is received, a supplemental mdr will be filed.
 
Event Description
It was reported to boston scientific corporation on (b)(6) 2018 that an ultraflex tracheobronchial uncovered distal release stent was to be used during a stent placement procedure performed on (b)(6) 2018.According to the complainant, during the procedure, the stent was able to be deployed inside the patient, however, the delivery system was unable to be removed.It was noted that the proximal end of the stent was on the delivery system and was not fully expanded.Reportedly, the delivery system could not be removed without pulling the stent out from the patient; the stent was moved back and forth during withdrawal of the device.The stent and delivery system were successfully removed, and the procedure was completed with a different device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAFLEX¿ TRACHEOBRONCHIAL
Type of Device
PROSTHESIS, TRACHEAL, EXPANDABLE
Manufacturer (Section D)
BOSTON SCIENTIFIC - MARLBOROUGH
100 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC - GALWAY
Manufacturer Contact
nancy cutino
100 boston scientific way
marlborough, MA 01752
5086834000
MDR Report Key7722665
MDR Text Key115111160
Report Number3005099803-2018-02387
Device Sequence Number1
Product Code JCT
Combination Product (y/n)N
PMA/PMN Number
K012883
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 07/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/13/2020
Device Model NumberM00564520
Device Catalogue Number6452
Device Lot Number22113819
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/04/2018
Initial Date FDA Received07/26/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/14/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-