• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN LP LLC NORTH HAVEN LAPRO-CLIP; LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN LP LLC NORTH HAVEN LAPRO-CLIP; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number 8886848813
Device Problems Device Slipped (1584); Firing Problem (4011)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 01/30/2017
Event Type  malfunction  
Manufacturer Narrative
Evaluation summary post market vigilance (pmv) led an evaluation of four devices.Visual examination of three cartridges noted that the clips were fully formed but not completely deployed from the cartridge.Since the clinical instrument was not received the loading units were loaded into a pmv representative instrument and fired; the pushers advanced properly and deployed the clips.The last clip body was received without the cartridge.The tracks were not inserted properly at the distal end of the clip body.Since the cartridge was not received a functional evaluation was not performed.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.Replication of the clip not fully deployed condition may occur if the trigger of the instrument is not fully actuated releasing the clip from the cartridge.Replication of the malformed clip condition may occur if the clip was not properly applied and mishandled during the clinical application.The root cause of the observed damage was misuse of the product which would have caused or contributed to the reported incident.If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during a laparoscopic cholecystectomy, the surgeon attempted to fire the clip applier but the clip was not placed into the jaws.When the clip applier was removed from the tissue, the clip also removed.This same event occurred several more times.No patient injury.Another device was used to complete the procedure.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAPRO-CLIP
Type of Device
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
MDR Report Key7722932
MDR Text Key115355900
Report Number1219930-2018-04108
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier10884521057661
UDI-Public10884521057661
Combination Product (y/n)N
PMA/PMN Number
K925602
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 10/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8886848813
Device Catalogue Number8886848813
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/04/2018
Initial Date Manufacturer Received 06/28/2018
Initial Date FDA Received07/26/2018
Supplement Dates Manufacturer Received08/08/2018
Supplement Dates FDA Received10/04/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-