Product event summary: the device was returned and analyzed.Visual inspection of the sheath showed that the stopcock was intact with no apparent issues.Also, the shaft was kinked 1.6 inch from the tip of the shaft.Air aspiration was reproduced during the pressure test when the test dilator was introduced through the sheath.The hemostatic valve was leaking and a torn valve disk was suspected.If information is provided in the future, a supplemental report will be issued.
|
It was reported that during a cryo ablation procedure, the sheath could not be smoothly operated.Additionally, upon inserting the balloon catheter in the sheath several times, air was continuously introduced, and when the sheath handle was operated, the curve shape was insufficient.It was also reported that the torque was difficult to be applied on the sheath.Therefore, the sheath was replaced which resolved the issues.The case was completed with cryo.No patient complications have been reported as a result of this event.
|