In the initial report, 7/8/2018 was listed as the report submission due date and 7/27/2018 is the correct report submission due date.Additionally, 7/12/2018 was incorrectly reported as the date the device was returned to the manufacturer.The thermocool® smart touch¿ electrophysiology catheter was returned on 6/4/2018.Corresponding fields of this report have been updated accordingly.Manufacture ref no: (b)(4).
|
Device evaluation summary: upon receipt, the device was visually inspected, and electrode #2 was found damaged with foreign material underneath.The catheter outer diameter was measured, and electrode #2 was found out specification due to the damage observed.Scanning electron microscope (sem) analysis was performed on the damaged area, and the results showed evidence of mechanical damage on the surface of the ring, as well as a sharp ring edge.It is possible that the damage was caused by contact with an unknown object.No other anomalies were observed.Fourier transform infrared spectroscopy (ftir) analysis was performed, and the results showed that the foreign material was primarily composed of an acrylonitrile-butadiene-styrene (abs) based material.The device history record (dhr) was reviewed, and no anomalies were found related to this complaint.In addition, the dhr review verifies that the device was manufactured in accordance with documented specifications and procedures.During the manufacturing process, all catheters are inspected for visual damage prior to packaging.On-line inspections and functional tests are in place to prevent catheters this type of damage from leaving the facility.The customer complaint was confirmed.The root cause of the foreign material underneath the electrode and the damage observed cannot be determined, since there is evidence that the device was manufactured in accordance with documented specification and procedures.Manufacturer¿s reference number: (b)(4).
|