Device Problem
Mechanical Problem (1384)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 07/11/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
|
Event Description
|
It was reported that a bd sero-fuge¿ 2001 centrifuge, 1 speed, 115v was spinning with the lid open.
|
|
Manufacturer Narrative
|
Due to an it issue beginning on 7/3/2018.Previously filed emdr's did not contain required fields.This supplemental emdr is filed to provide the following omitted fields: event attributed to: other.Device single use?: no.Device returned to manufacture: yes.
|
|
Event Description
|
It was reported that a bd sero-fuge 2001 centrifuge, 1 speed, 115v was spinning with the lid open.
|
|
Manufacturer Narrative
|
Investigation summary: the centrifuge was returned to bd for investigation.The observations from manufacturing engineering were: observations: 1.Visual observation of a lot of use and spills.2.Evidence of customer tempering/modifications; evidence of the solenoid or latch mechanism had been replaced.3.Motor control device (triac) was shorted from mt1 to mt2.Further investigation showed evidence of a dried liquid between these two pins on the back side of the pcb which supplied power to the motor causing it to spin whenever power was applied.The low impedance short most likely caused excessive heat to build in this area which led to the burning smell.As a note the circuit board is in close proximity of the housing in this area so any liquid that may have gotten into this area could get trapped.4.The short circuit was on the pcb assembly and not the component.5.The unit was plugged in and observed to spin with the lid open or closed however the lid latch operated correctly.6.This instrument was manufactured in sep 2009, and is approximately 9 years old.Investigation conclusion: the complaint of a burning smell and spinning with the lid open was confirmed.Based on the condition and the age of the centrifuge it is apparent that it was heavily used.There were previous repairs made to the instrument but did not led to the problem.The cause was most likely due to a liquid that caused a short on the motor control component.Capa # 226876.Trends and corrective action: the alert and action level for safety related complaints is 2 and 3 respectively.There were 3 complaints related to safety in the month of july 2018.As a result, action level was reached.Quality has initiated capa pr 226876 to address ¿running with lid open¿ issue, please refer to capa for further details.Quality will continue to monitor and investigate complaints relative to this trend.
|
|
Event Description
|
It was reported that a bd sero-fuge¿ 2001 centrifuge, 1 speed, 115v was spinning with the lid open.
|
|
Search Alerts/Recalls
|