• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN LP LLC NORTH HAVEN ENDO GIA; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN LP LLC NORTH HAVEN ENDO GIA; STAPLE, IMPLANTABLE Back to Search Results
Model Number EGIA45AVM
Device Problems Failure to Advance (2524); Failure to Fire (2610); Firing Problem (4011)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/13/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter during a gastrectomy, they connected the reload to the handle, then the surgeon tried to fire the device, however could not.The sales rep checked the cartridge in question and noted the staple pusher was slightly pushed and sled was slightly advanced.The procedure was completed with another device.There was no patient harm.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.Visual inspection of the returned product noted that the reload was partially fired with the interlock engaged.The jaws were open.Staple pushers were visible at the 3 cm cut line.Functionally the reload was loaded into a post market vigilance instrument, the interlock was over ridden, and the reload was applied to test media.All remaining staples were placed, and test media was cleanly transected.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of the partial fire condition with interlock engagement may occur when the firing handle has not been completely cycled.If the instrument return knobs are retracted at any point once the firing cycle has begun, the loading unit will engage into safety interlock and prevent further attempts to fire by ceasing the placement of staples and tissue transection, and prevent patient harm.The root cause of the observed damage was misuse of the product which caused or contributed to the reported condition.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO GIA
Type of Device
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mc dermott rd
north haven CT 06476
Manufacturer (Section G)
COVIDIEN LP LLC NORTH HAVEN
195 mc dermott rd
north haven CT 06476
Manufacturer Contact
lisa hernandez
60 middletown ave
north haven, CT 06473
2034925563
MDR Report Key7733387
MDR Text Key115647516
Report Number1219930-2018-04182
Device Sequence Number1
Product Code GDW
UDI-Device Identifier10884523003178
UDI-Public10884523003178
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K111825
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 12/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2023
Device Model NumberEGIA45AVM
Device Catalogue NumberEGIA45AVM
Device Lot NumberN8B0412KX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/10/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/13/2018
Initial Date FDA Received07/30/2018
Supplement Dates Manufacturer Received08/08/2018
11/17/2018
Supplement Dates FDA Received10/02/2018
12/11/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/11/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-