Model Number EGIA45AVM |
Device Problems
Failure to Advance (2524); Failure to Fire (2610); Firing Problem (4011)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 07/13/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter during a gastrectomy, they connected the reload to the handle, then the surgeon tried to fire the device, however could not.The sales rep checked the cartridge in question and noted the staple pusher was slightly pushed and sled was slightly advanced.The procedure was completed with another device.There was no patient harm.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.Visual inspection of the returned product noted that the reload was partially fired with the interlock engaged.The jaws were open.Staple pushers were visible at the 3 cm cut line.Functionally the reload was loaded into a post market vigilance instrument, the interlock was over ridden, and the reload was applied to test media.All remaining staples were placed, and test media was cleanly transected.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of the partial fire condition with interlock engagement may occur when the firing handle has not been completely cycled.If the instrument return knobs are retracted at any point once the firing cycle has begun, the loading unit will engage into safety interlock and prevent further attempts to fire by ceasing the placement of staples and tissue transection, and prevent patient harm.The root cause of the observed damage was misuse of the product which caused or contributed to the reported condition.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|