• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE MAHWAH CS300; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE MAHWAH CS300; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number N/A
Device Problem No Audible Alarm (1019)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
The production device history record (dhr) for this iabp unit is not required to be reviewed per sop 01-061 since the device manufacture date is greater than one year from the event date.A getinge service territory manager (stm) was dispatched to evaluate this unit.The stm confirmed customer issue in logs and was able to reproduce issue once upon arrival but could not duplicate issue after initial alarm.The stm completed condensation removal procedure, re-terminated internal vacuum pneumatic connections, and replaced crm module as a precaution.The stm then, completed calibration check, functional testing and safety check to factory specifications.The unit passes all functional checks was returned to the customer and cleared for clinical use.
 
Event Description
It was reported that the cs300 intra-aortic balloon pump (iabp) generated a low vacuum alarm.It is unknown under which circumstances this event occurred; however, there was no patient involvement and no adverse event was reported.
 
Event Description
It was reported that during setup, the cs300 intra-aortic balloon pump (iabp) generated a low vacuum alarm.There was no patient involvement and no adverse event was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE MAHWAH
1300 macarthur blvd.
mahwah NJ 07430
Manufacturer (Section G)
DATASCOPE MAHWAH
1300 macarthur blvd.
mahwah NJ 07430
Manufacturer Contact
1300 macarthur blvd.
mahwah, NJ 07430
MDR Report Key7735195
MDR Text Key115648311
Report Number2249723-2018-01275
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberN/A
Device Catalogue Number0998-00-3023-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 07/09/2018
Initial Date FDA Received07/31/2018
Supplement Dates Manufacturer Received08/16/2018
Supplement Dates FDA Received09/04/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/01/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-