Product complaint #: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Evaluation statement: the complaint device was received and inspected.The complaint can be confirmed.It was observed that the trocar and sleeve have become welded together.Visual inspection revealed numerous striations along the trocar and the sleeve shaft.The location of the trocar interlocking pins is superior to the mating portion of the sleeve.If the pins are not seated within these groves then the trocar is spinning within the guide sleeve causing enough heat and friction to cause the "welding" of the two parts.The root cause for this failure can be attributed to user technique.A non-conformance was identified for this lot number however the complaint condition is unrelated to the non-conformance.At this point, no corrective action is required and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi: (b)(4).
|