DENTSPLY LLC CAVITRON JET PLUS ULTRASONIC SCALER AND AIR POLISHING PROPHYLAXIS SYSTEM; SCALER, ULTRASONIC Back to Search Results
Model Number G132
Device Problems Restricted Flow rate (1248); Inaccurate Flow Rate (1249); Overheating of Device (1437); Temperature Problem (3022)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
There has been a previous report received where lack of water flow has caused an overheating insert.Since an overheating insert could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though results are not available as of this report.Evaluation results will be submitted as they become available.
 
Event Description
While using a g132 scaler, the water flow was inconsistent and the handpiece was heating up; no injury resulted.
 
Manufacturer Narrative
Hpc does not have proper water flow from debris built up, estimate new water system, cracked bracket board.Product is outside of useful life.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAVITRON JET PLUS ULTRASONIC SCALER AND AIR POLISHING PROPHYLAXIS SYSTEM
Type of Device
SCALER, ULTRASONIC
Manufacturer (Section D)
DENTSPLY LLC
1301 smile way
york PA 17404
MDR Report Key7743326
MDR Text Key116055714
Report Number2424472-2018-00111
Device Sequence Number1
Product Code ELC
Combination Product (y/n)N
PMA/PMN Number
K052334
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 10/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG132
Device Catalogue NumberG132
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/16/2018
Initial Date Manufacturer Received 07/03/2018
Initial Date FDA Received08/02/2018
Supplement Dates Manufacturer Received09/07/2018
Supplement Dates FDA Received10/10/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-