BD MEDICAL (BD WEST) MEDICAL SURGICAL BD POSIFLUSH¿ SALINE SYRINGE; SALINE FLUSH Back to Search Results
Catalog Number 306574
Device Problem Contamination /Decontamination Problem (2895)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/16/2018
Event Type  malfunction  
Manufacturer Narrative
Investigation: bd received two 5ml saline syringes from the reported lot, 7206569, for evaluation.Our quality engineer was able to visually inspect the returned samples and observed a brownish material embedded in the barrel and luer tip.Next, the engineer performed ftir analysis on the embedded foreign matter (fm) by removing a small piece of fm and testing it.The results indicated that the material was most likely a piece of valve material or polyisoprene along with chlorinated polyethylene.Based off of the results of the visual inspection and ftir the engineer was able to verify the reported issue of foreign matter.The fm was confirmed to have come from a valve used in the resin drying process that was deteriorated and inducing foreign matter.The source of the fm was removed and replaced with a new valve.There was documented embedded fm on one of the inspections performed during the production run for batch 7206569.The molding barrel accumulator, rail conveyors and the packaging area were cleared by scrapping all components and the syringes.Also a stat sampling was performed and passed with zero found.
 
Event Description
It was reported with the use of the bd posiflush¿ saline syringe there was an issue with foreign matter.There was no report of injury or further medical intervention.
 
Manufacturer Narrative
Correction for reason code for no evaluation and if other specify.The following field(s) have been updated.If follow up what type? yes, yes, no.
 
Event Description
It was reported with the use of the bd posiflush saline syringe there was an issue with foreign matter.There was no report of injury or further medical intervention.
 
Event Description
It was reported with the use of the bd posiflush¿ saline syringe there was an issue with foreign matter.There was no report of injury or further medical intervention.
 
Manufacturer Narrative
Correction: due to an it issue beginning on 7/3/2018, previously filed emdrs did not contain required fields.This supplemental emdr is filed to provide the following omitted fields: event attributed to: other.Device single use?: no.Device returned to manufacturer: yes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD POSIFLUSH¿ SALINE SYRINGE
Type of Device
SALINE FLUSH
Manufacturer (Section D)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
MDR Report Key7748884
MDR Text Key116051702
Report Number1911916-2018-00432
Device Sequence Number1
Product Code NGT
Combination Product (y/n)N
PMA/PMN Number
K011982
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other,user facility
Type of Report Initial,Followup,Followup
Report Date 08/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date07/31/2020
Device Catalogue Number306574
Device Lot Number7206569
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/23/2018
Initial Date Manufacturer Received 07/18/2018
Initial Date FDA Received08/03/2018
Supplement Dates Manufacturer Received07/18/2018
07/18/2018
Supplement Dates FDA Received08/06/2018
09/12/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-