Catalog Number 366076 |
Device Problem
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 07/16/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
The manufacturing location for this product is elitech.This site is an oem manufacturing site.Medical device expiration date: unknown.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Device manufacture date: unknown.
|
|
Event Description
|
It was reported that the roller on the bd¿ sedisystem esr instrument stopped.There was no report of exposure, injury, or medical intervention noted.
|
|
Event Description
|
It was reported that the roller on the bd¿ sedisystem esr instrument stopped.There was no report of exposure, injury, or medical intervention noted.
|
|
Manufacturer Narrative
|
Oem manufacturer: the manufacturing location for this product is (b)(4).This site is an oem manufacturing site.Therefore, bd corporate headquarters in (b)(4) has been listed and the (b)(4) fda registration number has been used for the manufacture report number.
|
|
Manufacturer Narrative
|
Investigation summary: a service work order indicates that the instrument was evaluated, repaired, restarted and a performance check showed no issues.Investigation conclusion: based on the service work order conducted, no additional issues were observed.Root cause description: the root cause was determined to be a printer driver (software) issue.The issue was corrected and no further errors reported.Rationale: complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.The bd business team regularly reviews the collected data for identification of emerging trends.Correction: due to an it issue beginning on 7/3/2018, previously filed emdrs did not contain required fields.This supplemental emdr is filed to provide the following omitted fields: event attributed to: other.Device single use?: no.Device returned to manufacture: yes.
|
|
Event Description
|
It was reported that the roller on the bd¿ sedisystem esr instrument stopped.There was no report of exposure, injury, or medical intervention noted.
|
|
Manufacturer Narrative
|
Correction: the bd sedi-15 instrument (catalog no.366076) was discontinued in january 2010 and is no longer distributed by bd.Additional information was received on 10/30/2018 from the bd technical support team who conducts a repair service on the device referenced in this report.It was determined that this device is not a bd sedi-15 catalog no.366076, but rather a non-bd printer.A processing error had occurred where the printer was logged into the service request system as the ¿bd sedi-15 catalog no.366076¿.As a result, the current complaint referencing the bd sedi-15 catalog no.366076 will be cancelled in our complaint system as it is not associated with our bd products.We have also taken necessary actions to prevent this processing error from recurring in the future.
|
|
Event Description
|
It was reported that the roller on the bd¿ sedisystem esr instrument stopped.There was no report of exposure, injury, or medical intervention noted.
|
|
Search Alerts/Recalls
|