• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN KANGAROO; PUMP, INFUSION, ENTERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN KANGAROO; PUMP, INFUSION, ENTERAL Back to Search Results
Model Number 383401
Device Problems No Display/Image (1183); Failure to Power Up (1476)
Patient Problem No Patient Involvement (2645)
Event Date 07/19/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
The customer reported the feeding pump had a blank screen.When their biomed department went to check the device, the unit would not power on.The event was discovered during testing.There is no reported patient involvement.After follow-up request for more information, the customer states all that was written from biomed was "unable to read the screen".The screen seemed intact, without defect but nothing was displayed on the lcd.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
An evaluation of the kangaroo pump was performed for a blank screen.Liquid ingress caused corrosion, and is the result of customer misuse.All device history records are reviewed for quality inspections and parameter compliance prior to releasing the product for shipment.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KANGAROO
Type of Device
PUMP, INFUSION, ENTERAL
Manufacturer (Section D)
COVIDIEN
2824 airwest blvd
plainfield IN 46168
Manufacturer (Section G)
COVIDIEN
2824 airwest blvd
plainfield IN 46168
Manufacturer Contact
edward almeida
15 hampshire street
mansfield, MA 02048
5084524151
MDR Report Key7753941
MDR Text Key116602921
Report Number3008361498-2018-00407
Device Sequence Number1
Product Code LZH
UDI-Device Identifier10884521144019
UDI-Public10884521144019
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 10/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number383401
Device Catalogue Number383401
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/26/2018
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 07/20/2018
Initial Date FDA Received08/06/2018
Supplement Dates Manufacturer Received08/08/2018
08/08/2018
Supplement Dates FDA Received10/02/2018
10/12/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-