• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LIMITED RESPIRATORY HUMIDIFIER; BTT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LIMITED RESPIRATORY HUMIDIFIER; BTT Back to Search Results
Model Number HC550JHU
Device Problems Adverse Event Without Identified Device or Use Problem (2993); Insufficient Information (3190)
Patient Problems No Consequences Or Impact To Patient (2199); Obstruction/Occlusion (2422)
Event Type  Injury  
Manufacturer Narrative
(b)(4).Device evaluation anticipated, but not yet begun.We will provide a follow up report upon completion of our investigation.
 
Event Description
A healthcare facility in the uk reported via a fisher & paykel healthcare (f&p) field representative on behalf of a homecare patient, that an hc550 respiratory humidifier was allegedly delivering insufficient humidity which caused discomfort, desaturation, plugging off and thickening secretions and resulted to the patient being admitted to the hospital for an emergency trachea change.The hc550 respiratory humidifier was being used with an intersurgical breathing circuit, a bivona flextend tracheostomy tube and a resmed astral 100 ventilator.The patient's condition was reported to be stable and the patient has since returned home.
 
Event Description
A healthcare facility in the uk reported via a fisher & paykel healthcare (f&p) field representative on behalf of a homecare patient, that an f&p hc550 respiratory humidifier was allegedly delivering insufficient humidity which caused discomfort, desaturation, plugging off and thickening secretions and resulted to the patient being admitted to the hospital for an emergency tracheostomy change.The patient's condition was reported to be stable and the patient has since returned home.
 
Manufacturer Narrative
(b)(4).Method: the complaint hc550 respiratory humidifier was returned to the f&p regional office in the uk where a trained f&p service technician conducted a functional check as per the product technical manual.Results: there was no fault found with the complaint hc550 upon evaluation; the device performed within specification.Conclusion: the f&p hc550 respiratory humidifier was being used with an intersurgical breathing circuit at the time of the incident which may have contributed to the reported lack of humidity.The user instructions that accompany the hc550 state that, "the use of breathing circuits, chambers or other accessories which are not approved by fisher & paykel healthcare may impair performance or compromise safety." the user has been recommended to use fisher & paykel healthcare breathing circuits to optimize the humidity being delivered to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRATORY HUMIDIFIER
Type of Device
BTT
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LIMITED
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
MDR Report Key7755320
MDR Text Key116207889
Report Number9611451-2018-00660
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
PMA/PMN Number
K132017
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,foreign,user facilit
Type of Report Initial,Followup
Report Date 07/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberHC550JHU
Device Catalogue NumberHC550JHU
Device Lot Number2100068906
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 07/12/2018
Initial Date FDA Received08/06/2018
Supplement Dates Manufacturer Received07/12/2018
Supplement Dates FDA Received09/09/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
BIVONA FLEXTEND TRACHEOSTOMY TUBE; BIVONA FLEXTEND TRACHEOSTOMY TUBE; INTERSURGICAL BREATHING CIRCUIT; INTERSURGICAL BREATHING CIRCUIT; RESMED ASTRAL 100 VENTILATOR; RESMED ASTRAL 100 VENTILATOR; BIVONA FLEXTEND TRACHEOSTOMY TUBE; INTERSURGICAL BREATHING CIRCUIT; RESMED ASTRAL 100 VENTILATOR
Patient Outcome(s) Hospitalization;
Patient Age24 MO
-
-