• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GUIDANT CRM CLONMEL IRELAND COGNIS; IMPLANTABLE CHF PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GUIDANT CRM CLONMEL IRELAND COGNIS; IMPLANTABLE CHF PULSE GENERATOR Back to Search Results
Model Number P108
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/13/2018
Event Type  Injury  
Manufacturer Narrative
Analysis of the returned product is currently ongoing.This report will be updated upon completion of the analysis.
 
Event Description
Boston scientific received information that this device recorded a code 1003 indicative of battery voltage too low for the projected remaining capacity.Boston scientific received information that a request was made to have data from this device analyzed.Data analysis showed the battery appeared to be depleting more quickly than expected.Device replacement was recommended.This device was explanted and successfully replaced.No additional adverse patient effects were reported.
 
Manufacturer Narrative
This device was thoroughly inspected and analyzed upon receipt at our quality assurance laboratory.Review of the device memory indicated that a low voltage alert, code 1003, was recorded.The battery voltage was lower than expected, but still supported full device function.Using historical daily battery voltage measurement data, engineers determined that this device was demonstrating behavior consistent with a high current condition associated with a compromised low voltage capacitor connected to the device's battery.Low voltage capacitors are used in the device's high voltage charging operation in order to facilitate fast charge times.Malfunction of these capacitors resulted in a high current drain, which was depleting this device's battery faster than normal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COGNIS
Type of Device
IMPLANTABLE CHF PULSE GENERATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
tim degroot
4100 hamline ave. n
st. paul, MN 
6515826168
MDR Report Key7755959
MDR Text Key116203207
Report Number2124215-2018-14101
Device Sequence Number1
Product Code NVZ
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 07/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date05/01/2013
Device Model NumberP108
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/24/2018
Was the Report Sent to FDA? Yes
Initial Date Manufacturer Received 06/27/2018
Initial Date FDA Received08/07/2018
Supplement Dates Manufacturer Received07/26/2018
Supplement Dates FDA Received10/08/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/04/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
P108
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
-
-