Model Number EGIAUSTND
Device Problems Failure to Advance (2524); Misfire (2532); Failure to Form Staple (2579); Firing Problem (4011)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/23/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during laparoscopic sleeve gastrectomy, while firing across stomach, the device had a poor staple formation, incomplete staple line distally and it also did not fire.The stapler fired approximately five millimeters before the knife blade stopped advancing to cut the tissue and locked out.The doctor then pulled back on the black release knob.The doctor fired an instrument to complete the transaction of the staple line.There was no patient injury.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO GIA ULTRA
Type of Device
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
Manufacturer (Section G)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
Manufacturer Contact
lisa hernandez
60 middletown ave
north haven, CT 06473
2034925563
MDR Report Key7757456
MDR Text Key116598410
Report Number2647580-2018-03926
Device Sequence Number1
Product Code GDW
UDI-Device Identifier10884523003543
UDI-Public10884523003543
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K111825
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEGIAUSTND
Device Catalogue NumberEGIAUSTND
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/23/2018
Initial Date FDA Received08/07/2018
Supplement Dates Manufacturer Received08/08/2018
Supplement Dates FDA Received10/03/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age50 YR
Patient Weight164
-
-