MEDTRONIC MED REL MEDTRONIC PUERTO RICO ENTERRA; INTESTINAL STIMULATOR Back to Search Results
Model Number 3116
Device Problem Therapy Delivered to Incorrect Body Area (1508)
Patient Problems Undesired Nerve Stimulation (1980); Pain (1994)
Event Date 12/30/2013
Event Type  Injury  
Manufacturer Narrative
Continuation of medical devices: product id 4351-35, serial# (b)(4), implanted: (b)(6) 2013, explanted: (b)(6) 2016, product type lead.Product id 4351-35, serial# (b)(4), implanted: (b)(6) 2013, explanted: (b)(6) 2016, product type lead.Information references the main component of the system.Other relevant device(s) are: product id: 4351-35, serial/lot #: (b)(4), ubd: 04-oct-2015, udi#: (b)(4); product id: 4351-35, serial/lot #: (b)(4), ubd: 07-nov-2015, udi#: (b)(4).(b)(4).Pertain to product id 4351-35 serial# (b)(4) product type lead and product id 4351-35 serial# (b)(4) product type lead.Occupation: non-healthcare professional.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a consumer regarding a patient with an implantable neurostimulator (ins).It was reported that since implant, the patient could feel the stimulation pulsation in their stomach, which caused pain to shoot to their left arm and shoulder.It was noted that the patient¿s healthcare provider (hcp) originally told the patient that the patient was just imagining it, but later the hcp told the patient that they started implanting the leads in the back of the stomach instead of the front because the front location was near a nerve and they found that stimulation delivered near that nerve would cause painful stimulation to the left arm area.The patient had the system replaced and the lead location revised from the front of the stomach to the rear; it was noted that the sensation stopped as soon as the hcp did this.No further complications were reported/anticipated.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTERRA
Type of Device
INTESTINAL STIMULATOR
Manufacturer (Section D)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key7763103
MDR Text Key116448574
Report Number3004209178-2018-17694
Device Sequence Number1
Product Code LNQ
UDI-Device Identifier00643169174993
UDI-Public00643169174993
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
H990014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2015
Device Model Number3116
Device Catalogue Number3116
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 07/11/2018
Initial Date FDA Received08/08/2018
Supplement Dates Manufacturer Received08/08/2018
Supplement Dates FDA Received10/04/2018
Date Device Manufactured09/06/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age37 YR
-
-