MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® LUBRI-SIL® ALL-SILICONE FOLEY CATHETER; FOLEY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® LUBRI-SIL® ALL-SILICONE FOLEY CATHETER; FOLEY CATHETER Back to Search Results
Model Number 175808
Device Problems Break (1069); Leak/Splash (1354); Material Puncture/Hole (1504)
Patient Problem Inflammation (1932)
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device was not returned.
 
Event Description
It was reported that the catheter had a pinhole at the joint between the catheter and the hub, as a result the catheter leaked.Allegedly, the patient experienced inflammation to the labial region.No medical intervention was reported.
 
Event Description
It was reported that the catheter had a pinhole at the joint between the catheter and the hub, as a result the catheter leaked.Allegedly, the patient experienced inflammation to the labial region.No medical intervention was reported.
 
Manufacturer Narrative
The reported event was confirmed as manufacturing-related.Visual evaluation of the returned sample noted one opened (no original packaging present), used silicone catheter present.Visual inspection of the catheter surface noted a small hole estimated to be 0.024" in diameter approximately 0.0290" from the junction where the shaft meets the bifurcation.The catheter balloon was inflated with 10ml of methylene blue solution (3 drops 1% aq methylene blue per 100ml of distilled water) and no leaks were observed along the inflation lumen.The catheter drainage funnel was then flushed with a methylene blue solution and the solution immediately began to stream from the hole in the shaft.The active length of the catheter balloon was measured (0.4005") and found to be within specification (0.40" +/- 0.1").The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use was found adequate and states the following: ¿lubri-sil® all-silicone foley catheter warning: after use, this product may be a potential biohazard.Handle and dispose of in accordance with applicable local, state, and federal laws and regulations.Warning: on catheter, do not use ointments or lubricants having petrolatum base.They will damage silicone and may cause the balloon to burst.Caution: federal (u.S.A.) law restricts this device to sale by or on the order of a physician.Sterile unless package is opened or damaged.Caution: do not aspirate urine through drainage funnel wall.Storage: store catheters at room temperature away from direct exposure to light, preferably in the original box.Single use only.Do not reuse.Do not resterilize.For urological use only.Valve type: use luer slip tip syringe.Do not use needle.Recommended inflation capacities: 3cc balloon: use 5cc sterile water, 5cc balloon: use 10cc sterile water, 30cc balloon: use 35cc sterile water, do not exceed recommended capacities.Note: aggressive traction, particularly in the presence of suturing, is not recommended for silicone foley catheters.To deflate catheter balloon: gently insert a luer slip tip syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.If you notice slow or no deflation, re-seat the syringe gently.Allow the balloon to deflate slowly on its own.Do not aspirate or manually accelerate the deflation of the balloon.If permitted by hospital protocol, the valve arm may be severed.If this fails, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.Visually inspect the product for any imperfections or surface deterioration prior to use.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® LUBRI-SIL® ALL-SILICONE FOLEY CATHETER
Type of Device
FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key7763571
MDR Text Key117017869
Report Number1018233-2018-03479
Device Sequence Number1
Product Code EZL
UDI-Device Identifier00801741034275
UDI-Public(01)00801741034275
Combination Product (y/n)N
PMA/PMN Number
K984084
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Type of Report Initial,Followup
Report Date 11/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/03/2022
Device Model Number175808
Device Catalogue Number175808
Device Lot NumberNGBU2291
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/16/2018
Initial Date Manufacturer Received 07/18/2018
Initial Date FDA Received08/08/2018
Supplement Dates Manufacturer Received11/08/2018
Supplement Dates FDA Received11/14/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age4 YR
-
-