• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HILL-ROM MEXICO ADVANTA 2 FRAME; A/C POWERED ADJUSTABLE HOSPITAL BED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HILL-ROM MEXICO ADVANTA 2 FRAME; A/C POWERED ADJUSTABLE HOSPITAL BED Back to Search Results
Model Number P1190A000067
Device Problem Self-Activation or Keying (1557)
Patient Problem No Patient Involvement (2645)
Event Date 07/24/2018
Event Type  malfunction  
Manufacturer Narrative
The hill-rom technician found the left hand siderail cable and membrane needed to be replaced.Per the hill-rom service manual the advanta¿ 2 should be subject to an effective maintenance program.This will help make sure of a long, operative life for the advanta¿ 2 bed.The preventative maintenance will help to reduce downtime due to excessive wear failures.An annual service of the bed is advised in order to maintain its characteristics and performance.Make sure the head and foot side rails are not bent or twisted.Make sure the latch mechanism operates correctly.An audible click must be heard.Remove the side rail cover, and make sure the mounting screws are tight.Inspect the cable routing for pinching, binding, and damage.Make sure all functions on the caregiver control operate correctly.Repair or replace the side rail as necessary.A search of the hill-rom maintenance records showed hill-rom performed preventative maintenance on this bed in (b)(6) 2018.It is unknown if the facility performed any other preventative maintenance on this bed.The technician replaced the left hand siderail cable and membrane to resolve the issue.Based on this information, no further action is required.
 
Event Description
Hill-rom received a report from the account stating the head of the bed self articulates (self runs).The bed was located at 375 at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANTA 2 FRAME
Type of Device
A/C POWERED ADJUSTABLE HOSPITAL BED
Manufacturer (Section D)
HILL-ROM MEXICO
ave.del telefono no. 200 col. huinala
apodaca, nuevo leon 66640
MX  66640
Manufacturer Contact
lori daulton
1069 state route 46 east
batesville, IN 47006
8129312280
MDR Report Key7764310
MDR Text Key116777422
Report Number3006697241-2018-00063
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 07/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberP1190A000067
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/24/2018
Initial Date FDA Received08/08/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/08/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-