• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER TRAUMA KIEL LAG SCREW, TI GAMMA3® Ø10.5X95MM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER TRAUMA KIEL LAG SCREW, TI GAMMA3® Ø10.5X95MM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Catalog Number 30600095S
Device Problem Failure to Osseointegrate (1863)
Patient Problem No Code Available (3191)
Event Date 04/20/2016
Event Type  Injury  
Manufacturer Narrative
It was noted that the device is not available for evaluation.If additional information is received, it will be provided in a supplemental report.Device disposition is unknown.
 
Event Description
The manufacturer became aware of adverse events from the norwegian hip fracture register.The title of this report is ¿gamma 3 in the norwegian hip fracture register¿ which is associated with the gamma 3 locking nail system, within that report, post-operative complications/ adverse events were reported, which occurred between 1-jan-2005 and 31-dec-2016.A review of the complaint handling database revealed that the events have not been reported previously to stryker, therefore 290 complaints were initiated for the revision surgeries retrospectively for different adverse events mentioned in the report.This product inquiry addresses revision surgery due to malunion.Re-osteosynthesis was performed.The 19 out of 290 cases.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAG SCREW, TI GAMMA3® Ø10.5X95MM
Type of Device
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
Manufacturer Contact
rose haas
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key7778701
MDR Text Key116997776
Report Number0009610622-2018-00736
Device Sequence Number1
Product Code HSB
UDI-Device Identifier04546540374936
UDI-Public04546540374936
Combination Product (y/n)N
Reporter Country CodeNO
PMA/PMN Number
K032244
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type literature
Reporter Occupation Other
Type of Report Initial
Report Date 08/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number30600095S
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 07/17/2018
Initial Date FDA Received08/13/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-