The customer contacted a siemens customer care center (ccc) and reported that reagent arm 1 probe and reagent lot were replaced on the dimension exl 200 instrument due to the discordant eco2 results.The customer reported that after replacing the reagent and probe, calibration and quality controls (qcs) were acceptable.The customer indicated that the reagent used in initial testing was compromised, and resolved the issue by replacing the reagent.A siemens customer service engineer (cse) was dispatched to the customer site to collect data for further analysis.Siemens further analyzed the instrument files and determined that qc was within range and there were no additional discordant results obtained on the instrument using the enzymatic carbonate flex reagent cartridge.The cause of the discordant, falsely elevated eco2 results is unknown.The instrument and reagent are operating within manufacturing specifications.No further evaluation of the device is required.
|
Discordant, falsely elevated enzymatic carbonate (eco2) results were obtained on three patient samples on a dimension exl 200 instrument using the enzymatic carbonate flex reagent cartridge lot eb9122.The customer replaced reagent arm 1 probe and repeated two samples, resulting in two discordant, falsely elevated eco2 results.The discordant results were not reported to the physician(s).Using a new reagent from lot eb9122, the samples were repeated on the same instrument, resulting lower.The 2nd repeat results were reported, as the correct results, to the physician(s).There are no reports of patient intervention or adverse health consequences due to the discordant, falsely elevated eco2 results.
|