Catalog Number K08-MTMF012B
Device Problem Packaging Problem (3007)
Patient Problem No Patient Involvement (2645)
Event Date 05/17/2018
Event Type  malfunction  
Manufacturer Narrative
The suspect device is expected to return for evaluation.A follow up will be submitted when the evaluation is complete.
 
Event Description
The distributor reported a defect in the packaging.This was identified during their initial inspection of received product.The device was not sent to a user facility.
 
Manufacturer Narrative
One device has been returned for evaluation.The product was examined visually.The complaint is confirmed.The root cause is attributed to rough handling during the manufacturing process.A search of the complaint database was performed and no similar complaints for this lot number were found.The device history record was reviewed, no exception documents were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERIT CUSTOM KIT
Type of Device
CUSTOM KIT
Manufacturer (Section D)
MERIT MEDICAL SYSTEMS INC.
1600 merit parkway
south jordan, UT 84095
Manufacturer (Section G)
MERIT MEDICAL SYSTEMS INC.
1600 merit parkway
south jordan, UT 84095
Manufacturer Contact
katie swenson cqe,cba,cqpa.
1600 merit parkway
south jordan, UT 84095
8012531600
MDR Report Key7786279
MDR Text Key117294337
Report Number1721504-2018-00080
Device Sequence Number1
Product Code DQO
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Expiration Date02/28/2021
Device Catalogue NumberK08-MTMF012B
Device Lot NumberH1303970
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/15/2018
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 07/17/2018
Initial Date FDA Received08/15/2018
Supplement Dates Manufacturer Received08/22/2018
Supplement Dates FDA Received08/31/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/05/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-