Model Number VICMO13.2
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Headache (1880); Intraocular Pressure Increased (1937); Vomiting (2144); No Code Available (3191)
Event Date 01/25/2018
Event Type  Injury  
Manufacturer Narrative
This product is manufactured in the u.S.But not marketed in the u.S.(b)(4).
 
Event Description
The reporter indicated the surgeon implanted a 13.2mm vicmo13.2 implantable collamer lens, -11.00 diopter, in the patient's right eye (od), on (b)(6) 2018.Two days after the surgery, the patient suffered from severe vomiting, headache, ac loss and tension of more than 40.The surgeon administered mannitol by iv and the symptoms decreased.After 1 week, the lens was explanted on (b)(6) 2018.A shorter lens was implanted and the problem was resolved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Type of Device
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
cynthia thai
1911 walker avenue
monrovia, CA 91016
6263037902
MDR Report Key7794679
MDR Text Key117563215
Report Number2023826-2018-01199
Device Sequence Number1
Product Code MTA
Combination Product (y/n)N
Reporter Country CodeEG
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Reporter Occupation Physician
Type of Report Initial
Report Date 07/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date10/31/2018
Device Model NumberVICMO13.2
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/24/2018
Initial Date FDA Received08/17/2018
Was Device Evaluated by Manufacturer? No
Date Device Manufactured11/12/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-