• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MANUFACTURING INC ACCELERATOR APS CENTRIFUGE MODULE; CENTRIFUGE MODULE FOR AUTOMATED SAMPLE HANDLING AND PROCESSING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MANUFACTURING INC ACCELERATOR APS CENTRIFUGE MODULE; CENTRIFUGE MODULE FOR AUTOMATED SAMPLE HANDLING AND PROCESSING SYSTEM Back to Search Results
Catalog Number 07L02-51
Device Problem Unintended Movement (3026)
Patient Problem No Code Available (3191)
Event Date 07/24/2018
Event Type  malfunction  
Manufacturer Narrative
All available patient information was included.Additional patient details are not available.An evaluation is in process.A final report will be submitted when the evaluation is complete.
 
Event Description
The customer observed the springs holding up the centrifuge lid failing on the aps centrifuge (ln 7l02).If the lid is left unsupported, the springs fail and the lid crashed down.The lid fell on the customers arm, but no injury occurred and no medical treatment was sought.
 
Manufacturer Narrative
Further investigation of the customer issue included a review of the complaint text, a search for similar complaints, review of instrument logs, a review of labeling and instrument service.No adverse trend was identified for the customer's issue.Labeling was reviewed and found to be adequate.The gas springs that hold the automated processing system (aps) centrifuge modules lid in the upright position failed and could not support the lids weight.The lid dropped onto the customers arm, no injury occurred.The field service representative decontaminated the module and replaced the worn gas spring with another identical component to resolve the issue.Based on all available information and abbott diagnostics' complaint investigation no product deficiency was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCELERATOR APS CENTRIFUGE MODULE
Type of Device
CENTRIFUGE MODULE FOR AUTOMATED SAMPLE HANDLING AND PROCESSING SYSTEM
Manufacturer (Section D)
ABBOTT MANUFACTURING INC
1921 hurd drive
irving TX 75038
MDR Report Key7797283
MDR Text Key117604365
Report Number1628664-2018-01502
Device Sequence Number1
Product Code JQP
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 12/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number07L02-51
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 07/24/2018
Initial Date FDA Received08/20/2018
Supplement Dates Manufacturer Received12/07/2018
Supplement Dates FDA Received12/11/2018
Is This a Reprocessed and Reused Single-Use Device? No