• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION MOBILE CHARGING SYSTEM; SCS CHARGING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION MOBILE CHARGING SYSTEM; SCS CHARGING SYSTEM Back to Search Results
Model Number 3711
Device Problem Temperature Problem (3022)
Patient Problem Burning Sensation (2146)
Event Type  Injury  
Manufacturer Narrative
This charger model was included in a field correction.Manufacturer's evaluation: corrective and preventive action (capa) investigation was performed.Pocket heating was confirmed.The investigation for the capa associated with heating while charging (pocket heating) concluded that the charger was capable of transferring energy to the ipg at a rate that would cause heating of the ipg and/or charging wand of sufficient elevated temperature to cause pain and burns.The heating while charging was determined to be exacerbated by off-axis charging of shallow implanted ipgs and that all chargers were capable of elevated heating.The manufacturer has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.The manufacturer defers to the patient's physician regarding medical history.
 
Event Description
Device 2 of 2: reference mfr.Report#: 1627487-2018-07660.It was reported the patient experienced heating at the ipg site while recharging.A company representative was scheduled to meet with the patient as the next course of action and evaluate the issue.
 
Event Description
Device 2 of 2: reference mfr.Report#: 1627487-2018-07660.Additional information received indicated no intervention is planned for this issue as it has improved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOBILE CHARGING SYSTEM
Type of Device
SCS CHARGING SYSTEM
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer Contact
jennifer shepard
6901 preston road
plano, TX 75024
9725264657
MDR Report Key7803256
MDR Text Key117828408
Report Number1627487-2018-07661
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup
Report Date 11/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/28/2011
Device Model Number3711
Device Lot Number3243523
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/25/2018
Initial Date FDA Received08/21/2018
Supplement Dates Manufacturer Received11/06/2018
Supplement Dates FDA Received11/29/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/08/2010
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number1627487/07/26/2012/001-C
Patient Sequence Number1
Patient Outcome(s) Other;
-
-