VYAIRE MEDICAL 3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV); VENTILATOR, HIGH FREQUENCY Back to Search Results
Model Number 3100 A
Device Problem Smoking (1585)
Patient Problem No Patient Involvement (2645)
Event Date 07/24/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The service technician traced the ground wire to the power module and found burnt traces.The service technician will replace the power module to address the reported issue.
 
Event Description
It was reported to vyaire medical that the service technician had performed a routine 7 year preventative maintenance on the ventilator and calibrations.The service technician stated that everything seemed fine.As he restarted the ventilator to finish calibrations, the ground wire seemed to be smoking and he was unable to go any further.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV)
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
VYAIRE MEDICAL
26125 north riverwoods blvd
mettawa IL 60045
Manufacturer (Section G)
VYAIRE MEDICAL INC.
1100 bird center drive
palm springs CA 92262
Manufacturer Contact
mindy faber
26125 north riverwoods blvd
mettawa, IL 60045
8727570116
MDR Report Key7806466
MDR Text Key117960795
Report Number2021710-2018-09229
Device Sequence Number1
Product Code LSZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 08/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3100 A
Device Catalogue Number768901
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/24/2018
Initial Date FDA Received08/22/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/19/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-