Additional: a5, d4, d10, h2, h6.Correction: b4, e4, g4, g7, h3, h10.Device history records (dhr): the device manufacturing quality records indicate that the released components met all requirements to perform as intended.No trend considering the following event is identified: instrument breakage.Event summary: it was reported that the dynesys torsion torque driver has seriously fractured when fastened with the blocker.This occured during surgery.Another torque driver with the same ref.Number was used.Review of received data: 3 pictures were received.All pictures show the fractured instrument.Devices analysis, visual examination: the returned instrument is broken into two pieces.The whole hex tip came apart.No further deformation can be seen on the instrument.The blue handle does not show any damages.Based on this visual examination the reported event can be confirmed.See pictures attached.Review of product documentation: the ifu outlines inspection and functional testing instructions for the end user and details specific steps to take if the instrument appears to be damaged or does not function as intended.- according to the cer the torsion torque driver is an alternative to the hexagonal screw driver.The torsion torque driver is inserted into the counter torque handle and is designed to tighten the set screws to a torque of 4 nm.Conclusion summary: it was reported that the instrument was broken during the surgery.In the complaint it is mentioned that the instrument was used between 50 and 60 times.As no surgical reports have been received, it cannot not be said how the instrument was used.It is not known if the instructions given in the surgical technique and ifu were followed for a correct instrument usage.It can be suspected that the instrument has been used intensely and exceeded the useful life.However, based on the available information, an exact root cause for the instrument breakage could not be determined.The need for corrective measures is not indicated and zimmer gmbh considers this case as closed.Zimmer biomet¿s reference number of this file is (b)(4).
|