Model Number 4415
Device Problem Infusion or Flow Problem (2964)
Patient Problem No Patient Involvement (2645)
Event Date 07/30/2018
Event Type  malfunction  
Manufacturer Narrative
The issue was discovered when the perfusionists were doing their normal cleaning and replacing of the fluid.As per the field service representative (fsr), it was verified that in heat mode, there was water trickling over the ice bank intermittently.He replaced the solenoid valve #1.The unit operated to manufacturer's specifications.The suspect part was returned to the manufacturer for further evaluation.
 
Event Description
It was reported that during the use of the device for a non-clinical activity, the flow rate was not where it should be.There was no patient involvement.
 
Manufacturer Narrative
The reported complaint was confirmed.During laboratory analysis, the product surveillance technician (pst) observed there to be no external damage of the solenoid valve.It appears that only the upper portion of the valve, which includes the solenoid, was replaced in the field as the lower valve portion is clean in comparison to the upper portion.Coil resistance measured 95 ohms, which is typical.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SARNS TEMP. CONTROL AND MONITOR SYSTEM (TCM II)
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS-SARNS TCM II HEATING AND COOLING
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key7812461
MDR Text Key118104991
Report Number1828100-2018-00439
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
PMA/PMN Number
K883603
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 09/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number4415
Device Catalogue Number4415
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/07/2018
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/02/2018
Initial Date FDA Received08/24/2018
Supplement Dates Manufacturer Received09/24/2018
Supplement Dates FDA Received09/24/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-