• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. HARMONIC ACE PLUS 7 W ADV HEMOSTASIS; INSTRUMENT, ULTRASONIC SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. HARMONIC ACE PLUS 7 W ADV HEMOSTASIS; INSTRUMENT, ULTRASONIC SURGICAL Back to Search Results
Catalog Number HARH45
Device Problem Failure to Cut (2587)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/01/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # p9299r.Investigation summary : the analysis results found that the device was received with the tissue pad detached and not returned but with evidence of body fluids and tissue pad material in the groove section of the clamp arm.The device was connected to a test hand piece and a gen11 and the device did activate during functional testing.The instrument was disassembled to inspect the internal components, no evidence was found that could have caused the reported activation issues.Based on the condition of the tissue pad, a probable cause of this damage is that the clamp of the device may have been closed and the instrument activated without tissue present.Care should be taken not to apply pressure between the instrument blade and tissue pad without having tissue between them.Keep the clamp arm open when back cutting or while the blade is active without tissue between the blade and tissue pad to avoid damage to the tissue pad.The resulting damage contributes to the removal of the pad from the clamp arm.The cleaning of the pad, not in accordance with the ifu, can also result in removal of the pad during use.The batch history record was reviewed and no defects, nc¿s or protocols related to the complaint, were found during the manufacturing process.
 
Event Description
It was reported that during an unknown procedure, the device couldn't cut any tissue during activation.There was no patient consequence reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HARMONIC ACE PLUS 7 W ADV HEMOSTASIS
Type of Device
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
milton garrett
475 calle c
guaynabo 00969
5133378865
MDR Report Key7813674
MDR Text Key118400975
Report Number3005075853-2018-12338
Device Sequence Number1
Product Code LFL
UDI-Device Identifier20705036014635
UDI-Public20705036014635
Combination Product (y/n)N
PMA/PMN Number
K132612
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 08/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2022
Device Catalogue NumberHARH45
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/23/2018
Initial Date Manufacturer Received 08/23/2018
Initial Date FDA Received08/24/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/05/2017
Is the Device Single Use? Yes
Type of Device Usage N
Patient Sequence Number1
-
-