The returned device, intended for use in treatment, was returned as an mdr for evaluation.There was a relationship found between the returned device and the reported incident.Visual inspection of the instrument show no manufacturing abnormalities.The front draw indicates a detached trap.The part was not returned with the device.The returned firstpass mini straight is a single used device and can only be limited functional tested.The needle was deployed into a dome foam model (om-3003) and performed as specified; however the trap at distal end is missing.The instrument is damaged and will not work any more as intended.The complaint was verified but the root cause could not be determined with certainty.There were no indications during manufacturing record review that would suggest that the device did not meet product specifications upon release into distribution.
|