STRYKER ORTHOPAEDICS-MAHWAH UNKNOWN TRIDENT 28MM 10° E LINER; HIP IMPLANT Back to Search Results
Catalog Number UNK_JR
Device Problems Degraded (1153); Unstable (1667); Device Damaged by Another Device (2915)
Patient Problem Injury (2348)
Event Date 07/31/2018
Event Type  Injury  
Manufacturer Narrative
An event regarding instability and range of motion issues (impingement) involving a unknown trident liner was reported.The event was not confirmed.Method & results: product evaluation and results: not performed as the device was not returned.Medical records received and evaluation: not performed as no medical records were returned for review.Product history review: not performed as the device lot details were not provided.Complaint history review: not performed as the device lot details were not provided.Conclusions: the exact cause of the event could not be determined because insufficient information was provided.Further information such as device details and return, pathology reports, pre- and post-operative x-rays and the primary operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time as no devices and insufficient information was received by stryker orthopaedics.If devices and / or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event Description
It was reported that patient's right hip was revised due to instability.Intra-operatively, poly wear was noted along the edge.Patient was revised to an adm/mdm liner construct and a new exeter head.Update: rep confirmed wear on the poly appeared to be caused by impingement with the stem.Stem was exeter but catalog and lot details are not available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN TRIDENT 28MM 10° E LINER
Type of Device
HIP IMPLANT
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
joann lavatelli
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key7823300
MDR Text Key118449528
Report Number0002249697-2018-02696
Device Sequence Number1
Product Code JDI
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation Physician
Type of Report Initial
Report Date 08/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberUNK_JR
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/31/2018
Initial Date FDA Received08/28/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age64 YR
Patient Weight76
-
-